The freedom driver was not supporting a patient.The reported issue is reported under two separate medical device reports: (1) freedom driver s/n (b)(4) (mfr report # 3003761017-2019-00014) and (2) freedom onboard battery s/n (b)(4) (mfr report # 3003761017-2019-00015).The customer, a syncardia certified hospital, reported that they were preparing a freedom driver s/n (b)(4) for a patient and the driver alarmed and then shut off.The customer also indicated the freedom onboard battery s/n (b)(4) may have caused the fault alarm.
|