(b)(4).Device evaluation by manufacturer: a field service engineer (fse) visited the customer's site to address the reported event.Fse verified error from the error log.Fse removed covers and thoroughly cleaned and lubricated y and z guide rails.Fse also removed needle and noted a slight bend and replaced the needle.Fse replaced the drvr2 board, inspected z motor electrical connection, reseated and tweaked sampler y alignment to center of patient sample tube.The customer is using new sample racks, fse ran over 90 patient samples and 7 diluted patient samples with no issue or errors.Instrument is operating as expected.Error reoccurred after several days, fse returned to customer's site to resolve issue.Fse verified from the printed error report that the 708 x1-axis error reoccurred.During evaluation, the fse found that error only occurs when the 10th sample in rack is introduced to upper rollers if sample tube is slightly lifted.Fse removed loader, cleaned and lubricated every guide rail, roller, pia sensor and belt for x1 and x2 movement.Fse replaced x1 motor, re-installed loader and performed alignments for x, y and z positions.Fse also ran sample loader (sl) rotate over 20times without any error, cleaned all rack bases that contact lower roller.Customer successfully ran quality control and 20 patient samples without errors and within acceptable ranges.No further action required by field service.The g8 instrument is functioning as expected.A 13-month complaint history review and service history review for similar complaints was performed for serial number (b)(4).There were two (2) similar complaints identified during the searched period, which includes this event.The g8 variant analysis mode operator's manual under chapter 6, troubleshooting states the following: error messages: when consulting with technical support about a problem, please note the error message and error number.In addition, if you follow the suggested solutions in this section and are still unable to resolve the error, or if you encounter an error message that is not noted, contact technical support.General error messages: with these errors, the assay stops and the analyzer immediately enters stand-by state.708 x1-axis error: an abnormality occurred in the operation error in the x1-axis.If this occurs, inspect the x1-axis.Then, execute sl rotate.The most probable cause of the reported event was due to faulty driver 2 board and x1 motor.
|
Tosoh bioscience, inc.Is submitting on behalf of the foreign manufacturer, tosoh corporation, per exemption number e2017013.Submission of this report does not constitute an admission that the importer or manufacturer's product caused or contributed to the event.H.3.Device evaluation by manufacturer: the driver 2 board was returned to tosoh instrument service center for investigation.Functional testing could not duplicate reported x1-axis error, but confirmed failure of board.Part will be discarded.The sample needle was returned to tosoh instrument service center for investigation.A visual inspection confirmed failed needle due to a bent tip.Part will be discarded.The most probable cause of the reported event was due to faulty driver 2 board and sampling needle.
|