(b)(4).Device evaluation by manufacturer: a field service engineer (fse) visited the customer to address the reported event.Fse confirmed complaint because the wash bottle was empty.Fse installed a new wash bottle and performed 4 drain flushes.Fse successfully completed qc and patient sample runs without any errors.No further action required by field service.The g8 instrument is functioning as expected.A 13-month complaint history review and service history review for similar complaints was performed for serial number (b)(4) through aware date.There were no similar complaints identified during the searched period.The g8 variant analysis mode operator's manual under chapter 6 troubleshooting states the following: 221 #### not detect (#### is the peak id): a specific peak (hemoglobin fraction) could not be detected.When this occurs repeatedly with some samples, the elution buffer may have become concentrated, resulting in unidentified peak detection on chromatograms.Never mix the buffers.When the error only occurs with specific samples, a hemoglobin variant may be present.The most probable cause of the reported event was due to operator error: customer was running with empty wash.
|
A customer reported they were getting error message 221 ##### not detect during quality control (qc) and patient sample runs on the g8 instrument.The customer stated total area (ta) = 0 and confirmed the column was recently replaced, column count (cc) = 011.Technical support specialist (tss) instructed the customer to check for any leaks and customer confirmed no leaks found.A field service engineer (fse) was dispatched to address the reported event, which resulted in delay in reporting of patient results for hemoglobin a1c (hba1c).There is no indication of any patient intervention or adverse health consequences due to the delay in reporting of patient results.
|