MAUDE Adverse Event Report: TOSOH CORPORATION TOSOH HLC-723G8 ANALYZER G8

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

TOSOH CORPORATION TOSOH HLC-723G8 ANALYZER G8 Back to Search Results
Model Number G8
Device Problem Mechanical Problem (1384)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 12/31/2018
Event Type  malfunction  
Manufacturer Narrative
(b)(4).Device evaluation by manufacturer: a field service engineer (fse) visited the customer's site to address the reported event.Fse was able to reproduce error when the instrument was powered up.Fse replaced the faulty sample needle, cleaned and lubricated sampling mechanism.Fse successfully completed quality control without any errors.No further action required by field service.The g8 instrument is functioning as expected.A 13 month complaint history review and service history review for similar complaints was performed for the serial number (b)(4) through aware date.There were no similar complaints identified during the search period.The g8 variant analysis mode operator's manual under chapter 5, maintenance procedures, under section 5.10 provides step-by-step instructions on the sampling needle assembly replacement.Chapter 6, troubleshooting states the following: 6.3 error messages: when consulting with technical support about a problem, please note the error message and error number.In addition, if you follow the suggested solutions in this section and are still unable to resolve the error, or if you encounter an error message that is not noted, contact technical support.General error messages: with these errors, the assay stops and the analyzer immediately enters stand-by state.710 z1-axis error: an abnormality occurred in the up and down movement of the sampling needle.If this occurs during a stat assay, check that the container setting (cup or tube) is correctly set.The error also occurs when the sample vial was not recognized as a primary tube, due to the disoriented sample sensor.The most probable cause of the reported event was due the faulty sample needle.
 
Event Description
A customer reported getting frequent 710 z1-axis error messages during sample run on the g8 instrument.A field service engineer (fse) was dispatched to address the reported event, which resulted in delay in reporting of patient results for hemoglobin a1c (hba1c).There is no indication of any patient intervention or adverse health consequences due to the delay in reporting of patient results.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TOSOH HLC-723G8 ANALYZER G8
Type of Device
G8
Manufacturer (Section D)
TOSOH CORPORATION
shiba-koen first building
3-8-2 shiba
minato-ku, tokyo, japan 10586 23
JA  1058623
Manufacturer (Section G)
TOSOH CORPORATION (MANUFACTURER)
shiba-koen first building
3-8-2 shiba
minato-ku, tokyo, japan 10586 23
JA   1058623
Manufacturer Contact
doria esquivel
6000 shoreline court
suite 101
south san francisco, CA 94080
6509368123
MDR Report Key8280119
MDR Text Key136147974
Report Number8031673-2018-05443
Device Sequence Number1
Product Code LCP
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K131580
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Reporter Occupation Other
Type of Report Initial
Report Date 01/25/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberG8
Device Catalogue Number021560
Was Device Available for Evaluation? Yes
Was the Report Sent to FDA? Yes
Date Report Sent to FDA01/25/2019
Distributor Facility Aware Date12/31/2018
Device Age6 YR
Event Location Outpatient Treatment Facility
Date Report to Manufacturer01/25/2019
Initial Date Manufacturer Received 12/31/2018
Initial Date FDA Received01/25/2019
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured04/01/2012
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-