(b)(4).Device evaluation by manufacturer: a field service engineer (fse) visited the customer's site to address the reported event.Fse was able to reproduce error when the instrument was powered up.Fse replaced the faulty sample needle, cleaned and lubricated sampling mechanism.Fse successfully completed quality control without any errors.No further action required by field service.The g8 instrument is functioning as expected.A 13 month complaint history review and service history review for similar complaints was performed for the serial number (b)(4) through aware date.There were no similar complaints identified during the search period.The g8 variant analysis mode operator's manual under chapter 5, maintenance procedures, under section 5.10 provides step-by-step instructions on the sampling needle assembly replacement.Chapter 6, troubleshooting states the following: 6.3 error messages: when consulting with technical support about a problem, please note the error message and error number.In addition, if you follow the suggested solutions in this section and are still unable to resolve the error, or if you encounter an error message that is not noted, contact technical support.General error messages: with these errors, the assay stops and the analyzer immediately enters stand-by state.710 z1-axis error: an abnormality occurred in the up and down movement of the sampling needle.If this occurs during a stat assay, check that the container setting (cup or tube) is correctly set.The error also occurs when the sample vial was not recognized as a primary tube, due to the disoriented sample sensor.The most probable cause of the reported event was due the faulty sample needle.
|