Model Number 460028MM
Device Problems Material Separation (1562); Insufficient Information (3190); Material Split, Cut or Torn (4008)
Patient Problems Mitral Regurgitation (1964); Regurgitation (2259); Fibrosis (3167)
Event Date 12/01/2018
Event Type  Injury  
Manufacturer Narrative
Post-operatively, the body undergoes various phases of the tissue response continuum resulting in the fibrous encapsulation of the ring.Local and systemic factors of the patient may play a role in the wound healing and inflammatory responses to biomaterials and implants.In this case, although ring dehiscence was noted the ring was explanted and was found to have fibrous material on the ring.The device was not returned for evaluation, as it was reported to be unavailable.Based on the available information, the root cause of the event was likely due to patient related factors.The device history record (dhr) was not reviewed as the reported event does not allege a malfunction that could be related to a manufacturing deficiency and/or one was not confirmed through investigation.Edwards will continue to review and monitor all events.Trends are monitored on a monthly basis and if action is required, appropriate investigation will be performed.
 
Event Description
Edwards lifesciences maintains an implant patient registry.This registry is a patient tracking mechanism for serialized edwards implantable devices (bioprosthetic heart valves and annuloplasty rings), rather than a true post-market surveillance registry.Through the registry, edwards is notified when these devices are implanted.In addition, patient and/or device status may be reported to the registry via the implantation data cards.The information is received from various sources (e.G.Surgeon, hospital and patient family members) and is not received in the form of a conventional "customer complaint." the information reported may or may not be related to the edwards device.In this case, it was reported via the implant patient registry that a 28mm mitral ring was explanted after an implant duration of 18 days due to severe regurgitation due to dehiscence and fibrous material on the ring.A 27mm valve was implanted in replacement.The patient was discharged to home in good condition on pod 5.
 
Manufacturer Narrative
Reference capa-20-00141.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
COSGROVE-EDWARDS ANNULOPLASTY SYSTEM
Type of Device
RING, ANNULOPLASTY
Manufacturer (Section D)
EDWARDS LIFESCIENCES
1 edwards way
irvine CA 92614
MDR Report Key8281927
MDR Text Key134229978
Report Number2015691-2019-00284
Device Sequence Number1
Product Code KRH
Combination Product (y/n)N
PMA/PMN Number
K923367
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional,user faci
Type of Report Initial,Followup
Report Date 01/04/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Expiration Date06/10/2023
Device Model Number460028MM
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 01/04/2019
Initial Date FDA Received01/28/2019
Supplement Dates Manufacturer Received07/23/2020
Supplement Dates FDA Received10/10/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Outcome(s) Hospitalization; Required Intervention;
Patient Age59 YR
-
-