U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BD INFUSION THERAPY SYSTEMS INC. S.A. DE C.V. BD CONNECTA¿ STOPCOCK WITH VALVE & EXTENSION TUBING

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BD INFUSION THERAPY SYSTEMS INC. S.A. DE C.V. BD CONNECTA¿ STOPCOCK WITH VALVE & EXTENSION TUBING Back to Search Results
Catalog Number 394971
Device Problem Leak/Splash (1354)
Patient Problem No Information (3190)
Event Date 01/18/2019
Event Type  malfunction  
Manufacturer Narrative
(b)(6).A device evaluation is anticipated, but has not yet begun.Upon completion of the investigation, a supplemental report will be filed.
 
Event Description
It was reported that two bd connecta¿ stopcock with valve & extension tubing were leaked from the drug administration cap while patient was hooked up to the product.
 
Manufacturer Narrative
Investigation: a device history review was conducted for lot number 8215502.Our records show that this is the only instance of this issue occurring in this production batch.According to the sampling plan applied for product performance, this lot was accepted and released without defects being noted during the final assembly or visual inspections.Additionally, our engineers were able to duplicate this event through the leakage testing of the submitted device.A subsequent review of our manufacturing line determined that the most likely root cause for this event is an abnormality in the equipment responsible for tubing assembly.Bd was able to confirm the defect with the provided sample, for leakage issue in leakage/air in injection valve that have been detected in others customer complaints, the team previously confirmed issues with this product lot 8037509 where leakage occurred at the valve port assy.Due to a bad tubing assembly.Engineering team assessed the assembly process finding a worn pin in station 5 that could cause the reported failure mode.This pin is in charge of assembling the gray tubing into the valve housing.
 
Event Description
It was reported that two bd connecta¿ stopcock with valve & extension tubing were leaked from the drug administration cap while patient was hooked up to the product.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BD CONNECTA¿ STOPCOCK WITH VALVE & EXTENSION TUBING
Type of Device
STOPCOCK
Manufacturer (Section D)
BD INFUSION THERAPY SYSTEMS INC. S.A. DE C.V.
periferico luis donaldo
colosio no. 579
nogales
MDR Report Key8283684
MDR Text Key134364174
Report Number9610847-2019-00111
Device Sequence Number1
Product Code FMG
Combination Product (y/n)N
PMA/PMN Number
N/A
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,other,user facility
Type of Report Initial,Followup
Report Date 04/03/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Other
Device Expiration Date07/31/2021
Device Catalogue Number394971
Device Lot Number8215502
Initial Date Manufacturer Received 01/18/2019
Initial Date FDA Received01/28/2019
Supplement Dates Manufacturer Received01/18/2019
Supplement Dates FDA Received04/04/2019
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Outcome(s) Other;
-
-