Catalog Number 306546 |
Device Problems
Crack (1135); Melted (1385); Material Twisted/Bent (2981); Sharp Edges (4013)
|
Patient Problem
No Information (3190)
|
Event Date 01/16/2019 |
Event Type
malfunction
|
Manufacturer Narrative
|
A device evaluation is anticipated, but has not yet begun.Upon completion of the investigation, a supplemental report will be filed.
|
|
Event Description
|
It was reported that a bd posiflush normal saline syringe was bent, melted, sharp, and partially cracked upon opening the package.
|
|
Manufacturer Narrative
|
Investigation summary: one sample was received for evaluation.This is the first complaint for the lot# 8242677 for the same defect or symptom.There was no documentation of issues for the complaint of batch #8242677 during this production run.A device history record review was completed with zero defects found.No quality notifications were written for this batch, nor for the associated assembly batches.Investigation conclusion: the luer tip is damaged/ deformed therefore failure mode is verified.No other damages were observed on the barrel, plunger rod or tip cap.Root cause description: if the tip cap cup somehow has a piece of plastic from a tip cap jam, then that piece will induce the damage to the luer.Rationale: complaints received for this device and reported condition will continue to be tracked and trended.Information will be captured on trend reports and monitored monthly.Our business team regularly reviews the collected data for identification of emerging trends.
|
|
Event Description
|
It was reported that a bd posiflush¿ normal saline syringe was bent, melted, sharp, and partially cracked upon opening the package.
|
|