According to the complaint received, a deviation for more than 50% was observed for the bilirubin parameter on abl90 flex plus analyzer placed in the delivery room.Two samples were measured at clinic, where the problem was reported ((b)(6)) and compared to measurements performed in the laboratory and measurements performed at another clinic ((b)(6)).The following bilirubin results were measured: sample 1 -(b)(6): 67 umol/l, abl90 flex analyzer.Sample 1 -(b)(6) (laboratory): 71 umol/l, abl90 flex analyzer.Sample 1 -(b)(6) (delivery room): 24 umol/l, abl90 flex plus analyzer.Sample 2 -(b)(6): 74 umol/l, abl90 flex analyzer.Sample 2 -(b)(6) (lab): 68 umol/l, abl90 flex analyzer.Sample 2 -(b)(6) (delivery): 17 umol/l, abl90 flex plus analyzer.Based on the above measurement it was reported that the abl90 flex plus analyzer in the delivery room demonstrated a low bias for more than <50% compared to the measurements performed on abl90 flex analyzers.There was no reports of serious injury or maltreatment in relation to this case.
|