MAUDE Adverse Event Report: DEPUY IRELAND 9616671 ATTUNE SPACER BLOCK; ATTUNE INSTRUMENTS : SPACER BLOCKS/TRIAL SHIMS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY IRELAND 9616671 ATTUNE SPACER BLOCK; ATTUNE INSTRUMENTS : SPACER BLOCKS/TRIAL SHIMS Back to Search Results
Catalog Number 254401014
Device Problems Break (1069); Component Missing (2306)
Patient Problem No Code Available (3191)
Event Date 01/08/2019
Event Type  malfunction  
Manufacturer Narrative
(b)(4).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event Description
It was reported that two of the springs on the extension/flexion spacer block broke off, preventing the attachment of shims.Surgical delay 1 minute.
 
Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.Product complaint #: (b)(4).Investigation summary: no device associated with this report was received for examination.Depuy synthes considers the investigation closed at this time.Should additional information be received, the information will be reviewed and the investigation will be re-opened as necessary.Device history lot: null.Device history batch: null.Device history review: null.If information is obtained that was not available for the initial medwatch , a follow-up medwatch, a follow-up medwatch will be filed as appropriate.
 
Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.Product complaint #: (b)(4).Investigation summary: the investigation confirmed two of the four balseal posts were broken and one balseal springs was missing.Depuy considers the investigation closed at this time.Should the additional information be received, the information will be reviewed and the investigation will be re-opened as necessary.If information is obtained that was not available for the initial medwatch, a follow-up medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ATTUNE SPACER BLOCK
Type of Device
ATTUNE INSTRUMENTS : SPACER BLOCKS/TRIAL SHIMS
Manufacturer (Section D)
DEPUY IRELAND 9616671
loughbeg, ringaskiddy co.
cork
EI 
MDR Report Key8285672
MDR Text Key134362247
Report Number1818910-2019-82659
Device Sequence Number1
Product Code HWT
UDI-Device Identifier10603295395379
UDI-Public10603295395379
Combination Product (y/n)N
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Type of Report Initial,Followup,Followup
Report Date 01/08/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number254401014
Device Lot NumberBFA19KM
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/26/2019
Initial Date Manufacturer Received 01/08/2019
Initial Date FDA Received01/29/2019
Supplement Dates Manufacturer Received02/12/2019
03/08/2019
Supplement Dates FDA Received02/13/2019
04/05/2019
Patient Sequence Number1
Patient Age70 YR
-
-