Catalog Number 502-03-56E
Device Problem Malposition of Device (2616)
Patient Problem Injury (2348)
Event Date 01/04/2019
Event Type  Injury  
Manufacturer Narrative
Reported event: an event regarding malposition involving a trident shell was reported.The event was not confirmed.Method & results: product evaluation and results: not performed as no product was returned for evaluation.Clinician review: no medical records were received for review with a clinical consultant.Product history review: indicated all devices were manufactured and accepted into final stock with no reported discrepancies.Complaint history review: there have been no other similar events for the lot referenced.Conclusions: the event itself could not be confirmed and exact cause of the event could not be determined because insufficient information was provided further information such as device return, pre and post operative x-rays, operative reports as well as patient history and follow up notes are required to complete the investigation for determining a root cause.No further investigation is required at this time.If additional information becomes available to indicate further evaluation is warranted, this record will be reopened.
 
Event Description
It was reported that the patient's right hip was revised due to a malpositioned shell and trunnionosis at the head/stem junction.Surgeon reported shell malposition occurred at implantation and that there are no allegations against the shell.No black tissue was noted in the patient.A 56mm tritanium shell with three screws, liner, and 36 +5 v40 lfit head were revised to a trident ii 64mm shell, mdm/adm liner construct, and 28 +4 biolox head.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PRIMARY TRITANIUM HEMI CLUSTER HOLE CUP 56MM
Type of Device
PROSTHESIS, KNEE PATELLOFEMOROTIBIAL, PARTIAL, SEMI-CONSTRAINED, CEMENTED, POLYM
Manufacturer (Section D)
STRYKER ORTHOPAEDICS-MAHWAH
325 corporate drive
mahwah NJ 07430
Manufacturer (Section G)
STRYKER ORTHOPAEDICS-MAHWAH
325 corporate drive
mahwah NJ 07430
Manufacturer Contact
joanne mahony
ida industrial estate
carrigtwohill NA
214532800
MDR Report Key8285953
MDR Text Key134370793
Report Number0002249697-2019-00366
Device Sequence Number1
Product Code NPJ
UDI-Device Identifier07613327040166
UDI-Public07613327040166
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K143085
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional,other
Reporter Occupation Physician
Type of Report Initial
Report Date 01/29/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Expiration Date12/31/2019
Device Catalogue Number502-03-56E
Device Lot NumberVJ46K1
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 01/04/2019
Initial Date FDA Received01/29/2019
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured12/22/2014
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Hospitalization; Required Intervention;
Patient Age58 YR
Patient Weight66
-
-