The single-use device was received for evaluation.Visual inspection did not identify any abnormalities that could have contributed to the reported condition.Leak testing and clear passage testing were performed with no issues noted.The device was found to flow normally with no defects or leaks observed.The reported condition was not verified.A batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot.Should additional relevant information become available, a supplemental report will be submitted.
|