(b)(4).Batch # p4t64j.Investigation summary: the analysis results found that the device was received with the tissue pad detached and not returned but with evidence of body fluids and tissue pad material in the groove section of the clamp arm.The device was connected to a test hand piece and a gen11 and the device did activate during functional testing.During functional testing, the device was activated for about 1 minute with no abnormal heat observed.The device was disassembled and no heat damage associated with the reported issue was found.Based on the condition of the tissue pad, a probable cause for this damage is that the clamp of the device may have been closed and the instrument activated without tissue present.Care should be taken not to apply pressure between the instrument blade and tissue pad without having tissue between them.Keep the clamp arm open when back cutting or while the blade is active without tissue between the blade and tissue pad to avoid damage to the tissue pad.The resulting damage contributes to the removal of the pad from the clamp arm.The cleaning of the pad, not in accordance with the ifu, can also result in removal of the pad during use.The batch history record was reviewed, and no defects, nc¿s, or protocols related to the complaint, were found during the manufacturing process.
|
It was reported that during a tumor resection of the parotid gland procedure, the shears presented a temperature increase and released the protective teflon from the tip of the device.When the teflon fell into the patient's cavity, it was removed by the surgeon immediately.The main reason the surgeon chose to replace the device was the overheating and teflon detachment since the patient makes use of a pacemaker and anticoagulant for cardiopathy.To preserve the patient's health, it was necessary to use another device to complete the procedure.According to report, there was nothing out of routine.The procedure was successfully completed and there was no change that would compromise the patient's health.
|