CORDIS CASHEL TRAPEASE PVCF FEM/JUG 55CM CSI; CATHETER, INTRAVASCULAR, DIAGNOSTIC Back to Search Results
Catalog Number 466P306AU
Device Problem Adverse Event Without Identified Device or Use Problem (2993)
Patient Problem Perforation of Vessels (2135)
Event Date 11/01/2018
Event Type  Injury  
Manufacturer Narrative
Occupation: other, senior counsel, litigation.Date of event: please note that the exact event date is unknown and the event date is the complaint awareness date.It was reported that a patient underwent placement of a trapease vena cava filter.The information provided indicated that the filter subsequently malfunctioned and caused injury and damage to the patient, including, but not limited to several filter struts perforating beyond the wall of vena cava.There is currently no additional information available for review.The product was not returned for analysis and the sterile lot number has not been provided; therefore, no device analysis nor device history record review could be performed.The trapease vena cava filter is indicated for use in the prevention of recurrent pulmonary embolism (pe) via percutaneous placement in the vena cava for patients in which anticoagulants are contraindicated, anticoagulant therapy for thromboembolic disease has failed, emergency treatment following massive pulmonary embolism where anticipated benefits of conventional therapy are reduced or for chronic, recurrent pulmonary embolism where anticoagulant therapy has failed, or is contraindicated.The purpose of a vena cava filter is to catch thrombus from the lower extremities as it travels along normal blood flow patterns up towards the heart.Vessel perforation is a known adverse event associated with implanting vena cava filters and is listed as such in the instructions for use (ifu).The ifu also notes vessel damage such as intimal tears and perforation as procedural and long-term complications related to ivc filters.With the limited information available and without the procedural films or post implant images to review the reported device perforation of the inferior vena cava could not be confirmed or further clarified.Given the limited information currently available for review, there is nothing to suggest that the reported events are related to the design and manufacturing process of the device; therefore, no corrective action will be taken.Should additional information become available, the file will be updated accordingly.Please note that this is the initial/final report for this product.
 
Event Description
As reported by the legal brief, the patient underwent placement of a trapease vena cava filter.The filter subsequently malfunctioned and caused injury and damage to the patient, including, but not limited to several filter struts perforating beyond the wall of vena cava.
 
Manufacturer Narrative
Additional information received per the patient profile form (ppf) states that the patient experienced perforation of filter strut(s) outside the inferior vena cava (ivc) and pain.The patient became aware of the reported event eight years and ten months after the index procedure the product was not returned for analysis.A review of the device history record revealed no anomalies during the manufacturing and inspection processes that can be associated with the reported complaint.Additional information is pending and will be submitted within 30 days of receipt.
 
Manufacturer Narrative
Additional information received per the patient profile form (ppf) states that the patient experienced perforation of filter strut(s) outside the inferior vena cava (ivc) and pain.The patient became aware of the reported event eight years and ten months after the index procedure.It was reported that a patient underwent placement of a trapease vena cava filter.The information provided indicated that the filter perforated the wall of the inferior vena cava and the patient has experienced pain as a result.There is currently no additional information available for review.The product was not returned for analysis.A review of the device history record revealed no anomalies during the manufacturing and inspection processes that can be associated with the reported complaint.The trapease vena cava filter is indicated for use in the prevention of recurrent pulmonary embolism (pe) via percutaneous placement in the vena cava for patients in which anticoagulants are contraindicated, anticoagulant therapy for thromboembolic disease has failed, emergency treatment following massive pulmonary embolism where anticipated benefits of conventional therapy are reduced or for chronic, recurrent pulmonary embolism where anticoagulant therapy has failed, or is contraindicated.The purpose of a vena cava filter is to catch thrombus from the lower extremities as it travels along normal blood flow patterns up towards the heart.Vessel perforation is a known adverse event associated with implanting vena cava filters and is listed as such in the instructions for use (ifu).The ifu also notes vessel damage such as intimal tears and perforation as procedural and long-term complications related to ivc filters.With the limited information available and without the procedural films or post implant images to review the reported device perforation of the inferior vena cava could not be confirmed or further clarified.Due to the nature of the complaint, the reported pain experienced by the patient could not be confirmed.Pain does not represent a device malfunction.Given the limited information currently available for review, there is nothing to suggest that the reported events are related to the design and manufacturing process of the device; therefore, no corrective action will be taken.Should additional information become available, the file will be updated accordingly.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TRAPEASE PVCF FEM/JUG 55CM CSI
Type of Device
CATHETER, INTRAVASCULAR, DIAGNOSTIC
Manufacturer (Section D)
CORDIS CASHEL
cahir road
cashel co. tipperary
EI 
MDR Report Key8290978
MDR Text Key134566188
Report Number1016427-2019-02470
Device Sequence Number1
Product Code DQO
Combination Product (y/n)N
PMA/PMN Number
K020316
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,other
Type of Report Initial,Followup,Followup
Report Date 05/06/2019
2 Devices were Involved in the Event: 1   2  
1 Patient was Involved in the Event
Date FDA Received01/30/2019
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Expiration Date08/31/2012
Device Catalogue Number466P306AU
Device Lot Number15024269
Was Device Available for Evaluation? No
Date Manufacturer Received04/06/2019
Patient Sequence Number1
Patient Outcome(s) Life Threatening;
-
-