Catalog Number HARH23
Device Problem Packaging Problem (3007)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 12/04/2018
Event Type  malfunction  
Manufacturer Narrative
(b)(4).Batch # r93x78.Investigation summary: the analysis results found that the harh23 device was returned inside its package; the package was returned partially open.Upon visual inspection, it was observed that the blister and the tyvek from the packaging were damaged; the blister was found to be broken at one of the sides of the package and the blister was noted to be wrinkly with one tear.In, addition pieces of the broken blister were still attached to the tyvek.Due to the damages found on the packaging, a possible cause for this condition is due to improper handling during transit or storage; it appears that the package hit a hard surface and this caused the reported event.All ees product is 100% inspected prior to release.The information you provided is compiled monitored and reviewed on a routine basis for any associated trends.The lot/batch history record was reviewed and no defects, ncr¿s or protocols related to the complaint, were found during the manufacturing process.
 
Event Description
It was reported that during the procedure, the doctor found the aseptic package was damaged.Primary package was damaged.Changed to another one to complete the procedure.There was no report on patient injury.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HARMONIC ACE PLUS 7 W ADV HEMOSTASIS
Type of Device
INSTRUMENT, ULTRASONIC SURGICAL
Manufacturer (Section D)
ETHICON ENDO-SURGERY, LLC.
475 calle c
guaynabo 00969
Manufacturer (Section G)
ETHICON ENDO-SURGERY, LLC.
475 calle c
guaynabo 00969
Manufacturer Contact
milton garrett
475 calle c
guaynabo 00969
5133378865
MDR Report Key8291237
MDR Text Key134591375
Report Number3005075853-2019-16281
Device Sequence Number1
Product Code LFL
UDI-Device Identifier10705036014676
UDI-Public10705036014676
Combination Product (y/n)N
PMA/PMN Number
K132612
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Other
Type of Report Initial
Report Date 01/17/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received01/30/2019
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date06/30/2023
Device Catalogue NumberHARH23
Device Lot NumberR93X78
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/02/2019
Date Manufacturer Received01/17/2019
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured07/19/2018
Is the Device Single Use? Yes
Type of Device Usage N
Patient Sequence Number1
-
-