MAUDE Adverse Event Report: ETHICON ENDO-SURGERY, LLC. HARMONIC ACE 36CM W ERG HANDLE; INSTRUMENT, ULTRASONIC SURGICAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ETHICON ENDO-SURGERY, LLC. HARMONIC ACE 36CM W ERG HANDLE; INSTRUMENT, ULTRASONIC SURGICAL Back to Search Results
Catalog Number ACE36E
Device Problem Failure to Cut (2587)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 11/30/2018
Event Type  malfunction  
Manufacturer Narrative
(b)(4).Batch # p9480u.Investigation summary: the device was returned with the distal tip of the blade broken off and not returned.The remaining blade portion was scratched and had evidence of contact with metal in or out of the operative field.This blade tip portion may have broken off of the device during transport to our analysis site.During functional testing on a gen11, an alert screen was displayed.A probable cause for the device to stop activating and for the gen11 to display an alert screen is blade damage.Due to the returned condition of the device, the reported event of "tissue effect issues", could not be confirmed.Probable causes of blade damage, including breakage, are external contact during pre-op or general use, blade contact with other devices, contact with staples or clips during the procedure or using any means other than the blade wrench to attach or detach the blade.Once minor blade damage has occurred, subsequent activation may increase the severity of the blade damage.This, in turn, can result in the device failing the pre-run test with the generator and displaying an alert screen.The alert screens that can result may include ¿tighten assembly,¿ ¿blade error detected,¿ or "relax pressure on blade," followed by a ¿replace instrument¿ screen later in the procedure.Continued usage of the damaged blade can result in a broken blade.The batch history record was reviewed and there were no defects, protocols or ncr(s) found during the manufacturing process related to this complaint.
 
Event Description
It was reported that during and unknown procedure, the scalpel hard was to cut and coagulate.Changed to another one to complete the procedure.There was no report on patient injury.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HARMONIC ACE 36CM W ERG HANDLE
Type of Device
INSTRUMENT, ULTRASONIC SURGICAL
Manufacturer (Section D)
ETHICON ENDO-SURGERY, LLC.
475 calle c
guaynabo 00969
Manufacturer (Section G)
ETHICON ENDO-SURGERY, LLC.
475 calle c
guaynabo 00969
Manufacturer Contact
milton garrett
475 calle c
guaynabo 00969
5133378865
MDR Report Key8291474
MDR Text Key134597866
Report Number3005075853-2019-16292
Device Sequence Number1
Product Code LFL
UDI-Device Identifier10705036000938
UDI-Public10705036000938
Combination Product (y/n)N
PMA/PMN Number
K042777
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Other
Type of Report Initial
Report Date 01/19/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date10/31/2022
Device Catalogue NumberACE36E
Device Lot NumberP9480U
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/18/2019
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 01/19/2019
Initial Date FDA Received01/30/2019
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured11/13/2017
Is the Device Single Use? Yes
Type of Device Usage Unknown
Patient Sequence Number1
Treatment
GENERATOR AND HANDPIECE
-
-