• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITH & NEPHEW ORTHOPAEDICS LTD ACETLR CUP HAP 58MM W/ IMPTR; PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/METAL, RESURFACING

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW ORTHOPAEDICS LTD ACETLR CUP HAP 58MM W/ IMPTR; PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/METAL, RESURFACING Back to Search Results
Catalog Number 74120158
Device Problems Adverse Event Without Identified Device or Use Problem (2993); Noise, Audible (3273)
Patient Problems Inflammation (1932); Pain (1994); Toxicity (2333); Test Result (2695)
Event Date 11/30/2018
Event Type  Injury  
Event Description
It was reported that a revision surgery from the right hip was performed due to elevated chromium and cobalt levels, metallosis with pain, metal on metal squeaking, chronic pseudomonas aeruginosa prostatitis.
 
Manufacturer Narrative
It was reported that right hip revision surgery was performed.During the revision, the acetabular cup & femoral head were removed.As of today, the implanted devices, all of which were used in treatment, and additional information has been requested for this complaint but has not become available.A review of the complaint history for the cup/head was performed using batch numbers in search of similar recurring reports for the products during their lifetimes.No other similar complaints were identified for the head + cup.In the absence of the actual devices, the production records were reviewed for the devices reportedly involved in this incident.Device history record review confirmed that all released parts met specifications applicable at the time of production.Review of the product ifu found adequate warnings and precautions in relation to the alleged failure modes.A risk management review was performed.No additional risks were identified as result of the reported event.The medical documents were reviewed.The reported pain, elevated metal levels, and stained fluid are consistent with an adverse reaction to metal debris; however, without the supporting lab results, imaging, and the analysis of the explanted components, the root cause of the reported reactions cannot be confirmed, and it cannot be concluded that the reported clinical reactions were associated with a mal-performance of the implant.Without return of the actual devices or further information we cannot further investigate or confirm the details supplied in this complaint, and our investigation remains inconclusive.If the products or additional information become available in the future, this case will be reopened.No preventative or corrective action has been initiated as a result of this investigation.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ACETLR CUP HAP 58MM W/ IMPTR
Type of Device
PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/METAL, RESURFACING
Manufacturer (Section D)
SMITH & NEPHEW ORTHOPAEDICS LTD
aurora house
spa park
leamington spa CV31 3HL
UK  CV31 3HL
MDR Report Key8292389
MDR Text Key134580492
Report Number3005975929-2019-00040
Device Sequence Number1
Product Code NXT
Combination Product (y/n)N
PMA/PMN Number
P040033
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type other
Type of Report Initial,Followup
Report Date 03/12/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Expiration Date08/31/2013
Device Catalogue Number74120158
Device Lot Number091990
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 01/22/2019
Initial Date FDA Received01/31/2019
Supplement Dates Manufacturer Received01/22/2019
Supplement Dates FDA Received03/13/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Treatment
UNKNOWN BHR DEVICE, UNKNOWN LOT; UNKNOWN BHR DEVICE, UNKNOWN LOT
Patient Outcome(s) Hospitalization; Required Intervention;
-
-