Model Number 0.850.0009
Device Problem Battery Problem (2885)
Patient Problem No Patient Involvement (2645)
Event Date 01/03/2019
Event Type  malfunction  
Manufacturer Narrative
The unit was returned, and an evaluation was conducted.Upon visual inspection, it appears that a thermal event occurred.The thermal event caused significant melting of the handset such that a complete root cause analysis cannot be performed.It is not possible to determine the exact sequence of events that led to the thermal event.There is evidence that a short circuit occurred in the battery pack assembly.The lower battery pack of handset fused to handset printed circuit board and melted plastic due to expansion and heating of lower battery pack, which also fused the handset to the charger.No apparent heating of upper battery pack.This concludes the investigation.
 
Event Description
It is reported that handset exploded while on the charging cradle.Damage to the counter top and other equipment in the office.There was no report of injury, patient or user involvement, or impact to patient care.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NOMAD PRO
Type of Device
EXTRAORAL SOURCE X-RAY SYSTEM
Manufacturer (Section D)
ARIBEX
11727 fruehauf drive
charlotte NC 28273
Manufacturer (Section G)
ARIBEX
11727 fruehauf drive
charlotte NC 28273
Manufacturer Contact
david waters
11727 fruehauf drive
charlotte, NC 28273
7045877297
MDR Report Key8293507
MDR Text Key135205125
Report Number1017522-2019-00003
Device Sequence Number1
Product Code EHD
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K081664
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional,user faci
Reporter Occupation Dentist
Type of Report Initial
Report Date 01/30/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number0.850.0009
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/15/2019
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 01/03/2019
Initial Date FDA Received01/31/2019
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured01/26/2010
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-