Model Number 9-ASD-032
Device Problem Unintended Movement (3026)
Patient Problem No Information (3190)
Event Date 11/21/2018
Event Type  Injury  
Manufacturer Narrative
An event of residual shunt and embolization was reported.The results of the investigation are inconclusive since the device was not returned for analysis.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed.Based on the information received, the cause of the reported incident could not be conclusively determined.Information from the field indicated that the device "slipped off" the ivc rim.This may be evidence that patient anatomy could have contributed to the event.
 
Event Description
On (b)(6) 2018, a 32mm amplatzer atrial septal occluder was implanted.During deployment, the device was stable across the atrial septum, but slipped off the ivc rim and a small residual shunt was observed.The user elected to retrieve the device tp mitigate possible embolization; however, during retrieval process, the device embolized to the rvot.The patient required surgical closure of the asd and simultaneously, the embolized device was removed, which were done successfully.The patient is reported to be stable and discharged.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AMPLATZER SEPTAL OCCLUDER
Type of Device
TRANSCATHETER SEPTAL OCCLUDER
Manufacturer (Section D)
AGA MEDICAL CORPORATION
5050 nathan lane north
plymouth MN 55442
Manufacturer (Section G)
AGA MEDICAL CORPORATION
5050 nathan lane north
plymouth MN 55442
Manufacturer Contact
stephanie o' sullivan
5050 nathan lane north
plymouth, MN 55442
6517565400
MDR Report Key8294914
MDR Text Key134716174
Report Number2135147-2019-00031
Device Sequence Number1
Product Code MLV
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P000039
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Physician
Type of Report Initial
Report Date 01/31/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Expiration Date12/31/2019
Device Model Number9-ASD-032
Device Catalogue Number9-ASD-032
Device Lot Number4928824
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 01/03/2019
Initial Date FDA Received01/31/2019
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured01/06/2015
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Hospitalization; Required Intervention;
Patient Age16 YR
-
-