It was reported that this cardiac resynchronization therapy defibrillator (crt-d) and right ventricular (rv) lead exhibited noise, oversensing, and pacing inhibition that resulted in approximately 3.5 seconds of asystole.Additionally, the patient's atrial arrhythmia was being oversensing by the rv lead and resulted in inappropriate anti-tachycardia pacing (atp) therapy being delivered.The rv lead also exhibited high out-of-range pacing impedance of greater than 2,000 ohms.The noise was able to be reproduced with isometrics, but the out-of-range impedance was not.The patient reported experiencing some light headedness that corresponds to when the issue was first noted.The rv lead was explanted and replaced.The crt-d remains in service.No additional adverse patient effects were reported.
|