No patient information provided as no patient was involved in this concern.A medtronic representative visited the site and completed system checkout.Testing revealed that upon start up out of the shipping crate, the system pendant showed "not ready" status for the x-ray generator.The system failed 2d and 3d imaging modalities.It was found that the x-ray generator was not ready upon initial startup of system.Internal generator inspection revealed that there was no power to any of the generator components.The power conversion enclosure tested, and no 400vdc present at j2 pins 1 & 3.The power conversion enclosure was replaced per and system functionality was restored.The power conversion enclosure was returned to the manufacturer for analysis.Analysis found that there was no power going to the x-ray generator.The enclosure power conversion failed bench test and transistor q30 failed; it was shorted.Analysis found that the reported event was related to an electrical issue.If information is provided in the future, a supplemental report will be issued.
|