• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BECTON DICKINSON AND COMPANY 10 ML BD POSIFLUSH¿ NORMAL SALINE SYRINGE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BECTON DICKINSON AND COMPANY 10 ML BD POSIFLUSH¿ NORMAL SALINE SYRINGE Back to Search Results
Catalog Number 306546
Device Problem Difficult to Flush (1251)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 01/14/2019
Event Type  malfunction  
Manufacturer Narrative
Date of event: unknown.The date received by manufacturer has been used for this field.Investigation summary: since no samples displaying the condition reported are available for examination, we were unable to fully investigate this incident.No root cause can be determined as no samples were received.We would be very interested in examining product that does not meet your expectations and our quality standards.Based on no sample, the investigation concluded, bd was not able to verify the indicated failure.A device history record review was completed with zero defects found.No quality notifications were written for this batch, nor for the associated assembly batches.It is noted however that this catalog number is not specified to be used with a syringe pump.Investigation conclusion: based on no sample, the investigation concluded, bd was not able to verify the indicated failure.Root cause description: undetermined however, this catalog number is not specified to be used with pumps.
 
Event Description
It was reported that a 10 ml bd posiflush¿ normal saline syringe was hard to prime.The nurse had to push hard to get it to flow, but then it would not flush a piv.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
10 ML BD POSIFLUSH¿ NORMAL SALINE SYRINGE
Type of Device
FLUSH
Manufacturer (Section D)
BECTON DICKINSON AND COMPANY
2153 12th avenue
columbus NE 68601
Manufacturer (Section G)
BECTON DICKINSON AND COMPANY
2153 12th avenue
columbus NE 68601
Manufacturer Contact
brett wilko
9450 south state street
sandy, UT 84070
8015652341
MDR Report Key8306648
MDR Text Key135149386
Report Number1911916-2019-00145
Device Sequence Number1
Product Code NGT
UDI-Device Identifier30382903065463
UDI-Public30382903065463
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K141311
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type other,user facility
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 01/17/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/05/2019
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date08/31/2021
Device Catalogue Number306546
Device Lot Number8242589
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received01/14/2019
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured08/30/2018
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Other;
-
-