A resolute onyx 4.00x15mm drug eluting stent was intended to be used during a procedure to treat a lesion in the mid rca, exhibiting mild tortuosity and severe calcification.No damage was noted to the packaging.No issues removing the device from the hoop.Device was inspected with no issues.Negative prep was performed with no issues.Device did not pass through a previously deployed stent.Resistance was encountered when advancing the device.It is reported that stent dislodged during delivery to the lesion.The device was not kinked or re-straightened during use.Multiple non-medtronic balloons were used to push the stent against the wall.The dislodged stent was then successfully covered with a resolute onyx 4.00x22mm.Post timi flow was 3, and lesion was 0% stenosis at the end of the procedure.No issues is reported for the second stent.
|