Additional information was requested but to date has not been received.The incident sample has been requested but to date has not been received for evaluation.If the sample is received, or if additional information pertinent to the incident is obtained a follow-up report will be submitted.As part of our manufacturing process, all device history records are reviewed and approved by quality, prior to release of product.
|
Additional information: d 4 unique identifier added.H 3: evaluation summary.The device history record (dhr) review of the reported lot number shows no discrepancies.Samples were received at the manufacturing plant for evaluation.All samples were received in the original package.Visual and functional evaluation was performed, and detachment was noticed at the connection between the cross spike and the tubing line.The reported failure mode will be treated as confirmed, related to the manufacturing/production process.The root cause and action plan will be documented through a formal corrective and preventative action (capa).This complaint will be closed with no further action and will be used for tracking and trending purposes.
|