• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ABBOTT POINT OF CARE I-STAT1 ANALYZER, IMMUNO READY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ABBOTT POINT OF CARE I-STAT1 ANALYZER, IMMUNO READY Back to Search Results
Catalog Number 04P75-36
Device Problem Electrical /Electronic Property Problem (1198)
Patient Problem No Patient Involvement (2645)
Event Date 01/29/2019
Event Type  malfunction  
Manufacturer Narrative
(b)(4).Apoc labeling will be evaluated during the investigation as pertaining to the event.
 
Event Description
On (b)(6) 2019, abbott point of care (apoc) was contacted by a customer who reported visible smoke coming from i-stat1 analyzer sn (b)(4) and would not power up after the event there was no additional information at the time of this report.The analyzer was replaced at no charge and returning for investigation.Apoc has determined that a component failure within the analyzer circuitry, may lead to the batteries becoming uncomfortably hot to touch in the area of the battery compartment when using a green non-fused battery carrier.However, the customer states that rechargeable batteries were being used at the time of the event.Therefore the analyzer is unlikely to become hot to touch.The product was replaced and returned for investigation.Based on the information available, there were no patient or user related injuries associated with this complaint.
 
Manufacturer Narrative
Apoc incident #: (b)(4).The investigation was completed on 03/01/2019.The incident was reviewed and was found to be identical in nature to an incident previously investigated.Because an identical scenario has already been investigated and this failure mode has been catalogued, no further investigation is necessary for this incident.No preventive/corrective action is required as the threshold has not been tripped and no product deficiency was found associated with the smoke emitted by the analyzer or with the inability of the analyzer to activate.Rather, this was a malfunction, which was attributed to the failure of the tantalum capacitor in the c13 location.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
I-STAT1 ANALYZER, IMMUNO READY
Type of Device
I-STAT1 ANALYZER
Manufacturer (Section D)
ABBOTT POINT OF CARE
400 college road
princeton NJ 08540 6607
MDR Report Key8312463
MDR Text Key137017725
Report Number2245578-2019-00021
Device Sequence Number1
Product Code CGA
UDI-Device Identifier00054749003754
UDI-Public00054749003754
Combination Product (y/n)N
PMA/PMN Number
K001387
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Type of Report Initial,Followup
Report Date 03/06/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number04P75-36
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/18/2019
Initial Date Manufacturer Received 01/29/2019
Initial Date FDA Received02/06/2019
Supplement Dates Manufacturer Received03/01/2019
Supplement Dates FDA Received03/06/2019
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-