The referenced electrode was not returned to olympus for evaluation.The cause of the reported event could not be confirmed.However, based on similar reported events the potential cause of the reported event could be attributed to the electrode coming in contact with a metallic object or the settings on the generator are too high.The instruction manual provides warning which states, use extreme caution when using electrosurgery in close proximity to or in direct contact with any metal objects.The working channel and operating sheaths of most rigid endoscopes are metal.Do not activate the instrument while any portion of the instrument tip is within the working channel or in contact with another metal object.Localized heating of the instrument and the adjacent metal object or working channel may result in damage to the contacting endoscope, and/or instrument tip.In addition, the instruction manual of the 744000 generator provides caution which states, examine all accessories and connections to the electrosurgical generator before use.Improper connection may result in arcs and sparks, accessory malfunction, or unintended surgical effects.First device report was submitted on mfr# 2951238-2019-00322.
|
Olympus was informed that during the middle of a transurethral resection of a prostate (turp) procedure, the charge nurse and surgeon reported that there was a spark observed inside the patient¿s prostate.The surgeon used a second similar device and the exact same thing occurred.The procedure was stopped before there was any actual fire.The patient was inspected; no harm was observed to the patient.The surgeon decided to stop using the generator and switched to monopolar energy using a different non-olympus generator to complete the procedure.In addition, no device component broke off and/or fell into the patient.There was no unexpected bleeding to the patient or delay in the procedure.There was no issue withdrawing the electrode from the patient.The generator was set to default settings for the pk button.The reported device was inspected prior to procedure with no anomalies observed.It is unknown if there was metal to metal contact.The reported device was discarded following the procedure.2 of 2.
|