• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MAQUET CARDIOPULMONARY AG HLM TUBING SET W/BIOLINE COATING; TUBING, PUMP, CARDIOPULMONARY BYPASS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MAQUET CARDIOPULMONARY AG HLM TUBING SET W/BIOLINE COATING; TUBING, PUMP, CARDIOPULMONARY BYPASS Back to Search Results
Model Number BEQ-HLS 7050 USA
Device Problem Noise, Audible (3273)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 01/25/2019
Event Type  malfunction  
Manufacturer Narrative
(b)(4).A follow-up medwatch will be submitted when additional information becomes available.(b)(4).
 
Event Description
According to the customer: during patient treatment a hls 7.0 was making noise and decoupled.The hls 7.0 was replaced.No known consequences to the patient.(b)(4).
 
Manufacturer Narrative
Maquet medical systems,usa(importer)submits this report on behalf of the legal manufacturer of the device maquet cardiopulmonary (b)(4).A follow-up medwatch will be submitted when additional information becomes available.Reference exemption # e2018002.Importer- (b)(4).The returned hls set was investigated in the laboratory of the manufacturer on 2019-02-13.A visual inspection was performed and clots are detected in the pump and on the blood inlet side.A water cycle with the cardiohelp was performed and at the beginning a noise was heard , after that the pump continued run without any noise.The pump was removed from the hls module.All magnets are available and mounted in the correct way.The bearing ball is also available.Under the bearing ball a air inclusion was detected.No other abnormalities were detected.Thus the failure could be confirmed.The most probable root for the reported failure will be investigated in (b)(4).Since the reported failure did not contribute to a death or serious injury no corrective action is needed.In addition at this time it cannot be concluded that this is a systemic error.No corrective action is needed.The data is also being handled through a designated maquet cardiopulmonary trending and applicable investigation process.Due to this no further investigation initiations will be completed at this time.
 
Event Description
Internal reference: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HLM TUBING SET W/BIOLINE COATING
Type of Device
TUBING, PUMP, CARDIOPULMONARY BYPASS
Manufacturer (Section D)
MAQUET CARDIOPULMONARY AG
rastatt
GM 
MDR Report Key8315706
MDR Text Key135717951
Report Number8010762-2019-00028
Device Sequence Number1
Product Code DWE
Combination Product (y/n)N
PMA/PMN Number
K080592
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Type of Report Initial,Followup
Report Date 06/18/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberBEQ-HLS 7050 USA
Device Catalogue Number701052794
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/08/2019
Initial Date Manufacturer Received 01/27/2019
Initial Date FDA Received02/07/2019
Supplement Dates Manufacturer Received06/18/2019
Supplement Dates FDA Received06/18/2019
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
-
-