ST. JUDE MEDICAL - NEUROMODULATION; SCS EXTENSION Back to Search Results
Model Number 3383
Device Problems Disconnection (1171); High impedance (1291)
Patient Problem Inadequate Pain Relief (2388)
Event Type  Injury  
Manufacturer Narrative
The results/method and conclusion codes along with investigation results will be provided in the final report.
 
Event Description
Device 3 of 4.Reference mfr report: 1627487-2019-01067, 1627487-2019-02160, 1627487-2019-02162.It was reported that the patient was experiencing high impedances and inefficient stimulation as a result of the extensions pulling out of the ipg header.The patient underwent surgery to explant and replace the extensions, which resolved the issue.
 
Manufacturer Narrative
The results of the investigation are inconclusive since the device was not returned for analysis.Based on the information received, the cause of the reported incident could not be conclusively determined.During processing of this complaint, attempts were made to obtain patient weight, but it could not be obtained.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Type of Device
SCS EXTENSION
Manufacturer (Section D)
ST. JUDE MEDICAL - NEUROMODULATION
6901 preston rd
plano TX 75024
Manufacturer (Section G)
ST. JUDE MEDICAL - NEUROMODULATION
6901 preston rd
plano TX 75024
Manufacturer Contact
andrea deitz
6901 preston road
plano, TX 75024
9723098000
MDR Report Key8324636
MDR Text Key135678238
Report Number1627487-2019-02161
Device Sequence Number1
Product Code GZB
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K960728
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 08/12/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/09/2019
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date12/13/2018
Device Model Number3383
Device Lot Number5765865
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received08/01/2019
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Other;
-
-