The cause of the non-reproducible advia centaur cp total human chorionic gonadotropin (thcg) results is being investigated by siemens.The instruction for use (ifu) under the interpretation of results section states the following: "results of this assay should always be interpreted in conjunction with patient's medical history, clinical presentation and other findings." the instruction for use (ifu) under the limitation section states the following: "all in vitro assays can generate erroneous results, both clinically false positive results (test results suggesting a condition that is absent) and clinically false negative results (test results failing to identify a condition that is present)." "there are many possible causes for these types of discordant results.Erroneous results may occur due to interference from identifiable serum constituents or patient-specific serum constituents." "if an aberrant or abnormal result, as defined by the laboratory protocol, occurs, laboratory personnel should first make certain that the system is performing and is operated and maintained in accordance with the product labeling.The user should then follow the laboratory protocol for advising the clinician of a result that appears to have deviated from the norms established by the laboratory." "test results alone are not diagnoses of medical conditions.For example, low titer elevations of hcg can occur in normal nonpregnant subjects.A physician's diagnosis involves evaluation of the test result in conjunction with, and in the context of, the patient's medical history, physical examination, and other test results sometimes in consultation with other medical experts.".
|
Mdr 1219913-2019-00015 was filed on february 11, 2019 reporting non-reproducible results with the thcg assay on the advia centaur cp on one sample.March 13, 2019 - additional information root cause was not identified.The customer reported no calibration or control performance issues although data was not provided.The integrity of the sample cannot be ruled out as the cause of the variability seen.Imprecision can be caused by pre-analytical issues including incomplete clotting or sample integrity in general.Customer has not reported further occurrences.No further investigation is required.
|