MAUDE Adverse Event Report: SMITH & NEPHEW, INC. GEN UNI ART INSRT XL-MG 8 MM; PROSTHESIS, KNEE, FEMOROTIBIAL, NON-CONSTRAINED, CEMENTED, METAL/POLYMER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW, INC. GEN UNI ART INSRT XL-MG 8 MM; PROSTHESIS, KNEE, FEMOROTIBIAL, NON-CONSTRAINED, CEMENTED, METAL/POLYMER Back to Search Results
Catalog Number 71248338
Device Problem Migration (4003)
Patient Problem Injury (2348)
Event Date 01/11/2019
Event Type  Injury  
Event Description
It was reported that a total knee revision surgery was performed due to prosthesis mobilization.
 
Manufacturer Narrative
The associated oxinium uni femoral component, genesis uni tibial baseplate and genesis uni articular insert were not returned for evaluation.The devices were manufactured in 2013.Device details were provided.Thus, our investigation including a review of complaint history for the listed parts revealed no prior complaints for the listed failure mode with the same batch number.A review of the manufacturing records for the listed batches did not reveal any deviation from the standard manufacturing processes.A clinical analysis noted that no relevant clinical supporting documents were provided to conduct a thorough analysis of the reported issue.Therefore, no clinical/medical assessment can be rendered at this time.Without the return of the actual product involved, our investigation of this report is inconclusive.No further investigation warranted for this complaint; however we will continue to monitor for future complaints and investigate as necessary.We consider this investigation closed.Should additional information be received, the complaint will be reopened.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GEN UNI ART INSRT XL-MG 8 MM
Type of Device
PROSTHESIS, KNEE, FEMOROTIBIAL, NON-CONSTRAINED, CEMENTED, METAL/POLYMER
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks road
memphis TN 38116
MDR Report Key8328744
MDR Text Key135852191
Report Number1020279-2019-00521
Device Sequence Number1
Product Code HSX
UDI-Device Identifier03596010500731
UDI-Public03596010500731
Combination Product (y/n)N
PMA/PMN Number
K912735
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,health professional,u
Type of Report Initial,Followup
Report Date 04/02/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/11/2019
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Expiration Date02/24/2023
Device Catalogue Number71248338
Device Lot Number13BM15187
Date Manufacturer Received01/21/2019
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Outcome(s) Hospitalization; Required Intervention;
-
-