• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BECTON, DICKINSON AND CO. BD POSIFLUSH¿ XS PRE-FILLED FLUSH SYRINGE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BECTON, DICKINSON AND CO. BD POSIFLUSH¿ XS PRE-FILLED FLUSH SYRINGE Back to Search Results
Catalog Number 306572
Device Problem Adverse Event Without Identified Device or Use Problem (2993)
Patient Problem No Code Available (3191)
Event Date 01/11/2019
Event Type  Injury  
Manufacturer Narrative
(b)(6).A device evaluation is anticipated, but has not yet begun.Upon completion of the investigation and/or device history review, a supplemental report will be filed.
 
Event Description
It was reported that after use of the bd posiflush¿ xs pre-filled flush syringe patient called to advise feeling unwell.Patient advised to attend hospital re: potential risk of sepsis.
 
Manufacturer Narrative
Investigation: a dhr was performed and the non-conformances were reviewed for this batch, and there was no record of non-conformance which could contribute to the complaint verbatim reported by the customer.No sample or photo was returned.Based on the information provided, it is more probable than not that the symptoms described may be an allergic reaction; however, it is highly improbable that this reaction was produced by the normal saline in the bd posiflush product.During the period 2017-2019, there are no other adverse customer complaint trends for the complaint category of allergic reaction, apart from fresenius kabi uk.There are no known allergies to normal saline, either topical or within the body.It is most likely that whatever fluid was in the iv line prior to flushing contained inadequately flushed medication or glucose from previous treatments.
 
Event Description
It was reported that after use of the bd posiflush¿ xs pre-filled flush syringe patient called to advise feeling unwell.Patient advised to attend hospital re: potential risk of sepsis.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BD POSIFLUSH¿ XS PRE-FILLED FLUSH SYRINGE
Type of Device
SYRINGE
Manufacturer (Section D)
BECTON, DICKINSON AND CO.
donore road
drogheda
MDR Report Key8331818
MDR Text Key135907806
Report Number9616657-2019-00122
Device Sequence Number1
Product Code NGT
Combination Product (y/n)N
PMA/PMN Number
N/A
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type distributor,foreign,other
Type of Report Initial,Followup
Report Date 02/27/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/12/2019
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Other
Device Expiration Date08/31/2021
Device Catalogue Number306572
Device Lot Number8242797
Date Manufacturer Received01/21/2019
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
-
-