Device evaluation the everflex entrust stent delivery system, (sds), was received for evaluation loosely coiled with the handle disassembled within a large red biohazard bag.No ancillary devices from the procedure were received for evaluation.The handle was received split into its two halves, thumbwheel and pulley loose along with the red safety tab and tube.The catheter is coiled, has several kinks along its length, and the blue isolation sheath/strain relief pulled proximally until it is touching the guidewire lumen proximal hub.The distal ends of the two handles halves were examined: no unusual wear marks from the pull cable were present.This indicates that the pull cable was properly routed.The proximal hub on the guidewire lumen was cut-off to allow further examination of the guidewire lumen.Once the hub was removed the isolation sheath was slid off the guidewire lumen.The guidewire lumen was placed in its groove in the handle a series of accordion folds/kinks in the guidewire lumen were noted just distal of the handle¿s safety tab cavity.The accordion folds/kinks are an indicator that the guidewire used during the procedure did not provide enough support.A kink in the guidewire lumen was noted just proximal of the proximal end of the silver outer sheath.This kink most likely happened post-procedure given how the device was packaged for return.The pull cable appears to have broken proximal to the outer sheath.The distal end of the catheter¿s outer sheath was examined: the distal end appears to be ovalized.Backlighting indicated the presence of sanguine residue within the outer sheath.If information is provided in the future, a supplemental report will be issued.
|