MAUDE Adverse Event Report: MEDTRONIC PUERTO RICO OPERATIONS CO. ACTIVA; STIMULATOR, ELECTRICAL, IMPLANTED, FOR PARKINSONIAN TREMOR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC PUERTO RICO OPERATIONS CO. ACTIVA; STIMULATOR, ELECTRICAL, IMPLANTED, FOR PARKINSONIAN TREMOR Back to Search Results
Model Number 37603
Device Problems Failure to Interrogate (1332); No Device Output (1435); Insufficient Information (3190)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 01/18/2019
Event Type  malfunction  
Manufacturer Narrative
Concomitant medical products: product id: 37092, serial #: unknown, product type: accessory.Other relevant device(s) are: product id: 37092, serial/lot #: unknown.If information is provided in the future, a supplemental report will be issued.
 
Event Description
Information was received from a patient regarding their implantable neurostimulator (ins) for essential tremor and movement disorders.It was reported the patient programmer was not working.The patient programmer was not powering on and they clarified there was a poor communication screen with the antenna plugged in.They turn the ins off at night and turn it back on in the morning, lately, they cannot do that.The patient changed the patient programmer four times and it did not help in resolving the issue.The poor communication has not resolved.When the patient disconnected the antenna from the patient programmer they were able to sync with the ins.They then tried to use the patient programmer again and they were seeing the poor communication screen.No patient symptoms or further complications were reported as a result of this event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ACTIVA
Type of Device
STIMULATOR, ELECTRICAL, IMPLANTED, FOR PARKINSONIAN TREMOR
Manufacturer (Section D)
MEDTRONIC PUERTO RICO OPERATIONS CO.
road 31, km. 24, hm 4
ceiba norte industrial park
juncos PR 00777
Manufacturer (Section G)
MEDTRONIC PUERTO RICO OPERATIONS CO.
road 31, km. 24, hm 4
ceiba norte industrial park
juncos PR 00777
Manufacturer Contact
lisa woodward clark
7000 central avenue ne rcw215
minneapolis, MN 55432
7635263920
MDR Report Key8335349
MDR Text Key137558062
Report Number3004209178-2019-03047
Device Sequence Number1
Product Code MHY
UDI-Device Identifier00613994761071
UDI-Public00613994761071
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P960009
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type consumer
Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date 02/13/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/13/2019
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date12/14/2017
Device Model Number37603
Device Catalogue Number37603
Was Device Available for Evaluation? No
Date Manufacturer Received01/22/2019
Date Device Manufactured06/15/2016
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age80 YR
-
-