(b)(4).The distributor field service engineer (fse) arrived at the site to address the reported event.Fse checked the software parameters and found many of them to be set incorrectly.Next, fse reinstalled the 5.24 software version to set all parameters to default.No further issues were noted.No further action was required.A 13 month complaint history review and service history review for similar complaints was performed for the serial number (b)(4) from 06nov2017 through aware date (b)(4) 2018.There were no similar complaints identified during the search period.The g8 operator's manual under chapter 6, troubleshooting, states the following: 6.3 error messages: when consulting with technical support about a problem, please note the error message and error number.In addition, if you follow the suggested solutions in this section and are still unable to resolve the error, or if you encounter an error message that is not noted, contact technical support.General error messages: with these errors, the assay stops and the analyzer immediately enters stand-by state.707 syringe-s error: explanation: operation error in syringes.Countermeasure: inspect x1-axis.Inspect syringe-s.Execute smp.Reset.The most probable cause of the reported event was due to incorrect parameter settings.
|