Model Number G38483
Device Problem Adverse Event Without Identified Device or Use Problem (2993)
Patient Problems Hypovolemia (2243); Disability (2371)
Event Date 10/02/2018
Event Type  Injury  
Manufacturer Narrative
Pma/510(k) # = p100022/s026.(b)(4).Device evaluation: the zisv6-35-125-6-140-ptx device of lot number c1525770 involved in this complaint was implanted in the patient, and was not available for evaluation.With the information provided, a document based investigation was conducted.There was 5 cook stents involved in this event.Reference pr252461 (2 stents), pr253168 (1 stent), pr253170 (1 stent), pr253171 (1 stent).Document review: prior to distribution all zilver zisv6 devices are subject to visual inspection and functional checks to ensure device integrity.A review of the relevant manufacturing records (c1525770) revealed no discrepancies that could have contributed to this complaint.Based on the information available to date, there is no evidence to suggest that there are any manufacturing issues associated with lot number c1525770 there is no evidence to suggest that the customer did not follow the instructions for use.Root cause review: this event of the patient undergoing amputation post stent placement was likely not stent related.Summary: there is no evidence to suggest that this incident did not occur.Therefore, the complaint is confirmed based on customer testimony.The patient did undergo amputation which was likely not stent related.Complaints of this nature will continue to be monitored for potential emerging trends.
 
Event Description
On (b)(6) 2018 an arteriogram of the lower leg with balloon angioplasty and stent placement was performed and 5 cook stents were placed.No flow was noted, and the physician indicated they did not believe it was due to the devices.On (b)(6) 2019 the dm followed up to check on the patient, and the physician informed the dm that the patient had to have amputation (unknown location).Patient is stable and healing well.Update per dm on (b)(6) 2019: "the wire used in the case was a terumo stiff straight glide wire.I have not been able to get imaging of the stents.After stent deployment there was no flow, i asked the physician if they thought it was related to the stents and they said no.".
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ZILVER PTX 35 DRUG-ELUTING STENT
Type of Device
NIU STENT, SUPERFICIAL FEMORAL ARTERY, DRUG-ELUTING
Manufacturer (Section D)
COOK IRELAND LTD
o halloran road
limerick
Manufacturer (Section G)
COOK IRELAND LTD
o halloran road
national technology park
limerick
Manufacturer Contact
heather ryan
o halloran road
national technology park
limerick 
061334440
MDR Report Key8339304
MDR Text Key136139837
Report Number3001845648-2019-00059
Device Sequence Number1
Product Code NIU
UDI-Device Identifier10827002384832
UDI-Public(01)10827002384832(17)200420(10)C1525770
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Physician
Type of Report Initial
Report Date 01/18/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Lay User/Patient
Device Expiration Date04/20/2020
Device Model NumberG38483
Device Catalogue NumberZISV6-35-125-6-140-PTX
Device Lot NumberC1525770
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Distributor Facility Aware Date01/18/2019
Event Location Hospital
Initial Date Manufacturer Received 01/22/2019
Initial Date FDA Received02/14/2019
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured07/31/2018
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Disability;
-
-