The reported event was unconfirmed.The evaluation found that the color of the lubricating jelly was clear and transparent as per specification.There were no abnormalities observed on the returned lubricating jelly.The device history record was reviewed and found nothing that could have caused or contributed to the reported event.The instructions for use were found adequate and state the following: "[contraindications]
1.Method for use:
(1) do not reuse.(2) do not resterilize.(3) this device contains 10% povidone-iodine.For patients with past history of allergic hypersensitivity to povidone-iodine or iodine, consider using alternative disinfectants.(4) be careful that the catheter is not exposed to ointments, contrast medium or oil-based lubricants (including vegetable oils such as olive oil, mineral oils such as white petrolatum and animal oils).[they may damage the device and may burst balloon.]
(5) do not hold the device with forceps, etc.Avoid contact with any blades or sharp-edged instruments.[catheter damage may cause balloon rupture and accidental balloon removal or failure to deflate or remove the balloon.]
2.Applicable patients
(1) patients who are or have been allergic to natural rubber latex
(2) patients with known allergy to silver coated catheter
[directions for use]
1.Method of use
the device is intended for single use only and is not reusable.(5) lubricate the distal end of the catheter with water-soluble lubricant packaged in the tray.".
|