• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ETHICON ENDO-SURGERY, LLC. LIGACLIP*ENDO ROTATING MCA; CLIP, IMPLANTABLE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ETHICON ENDO-SURGERY, LLC. LIGACLIP*ENDO ROTATING MCA; CLIP, IMPLANTABLE Back to Search Results
Catalog Number ER320
Device Problem Material Puncture/Hole (1504)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Type  malfunction  
Manufacturer Narrative
(b)(4).Date of event: only event year known: 2019.Batch # unk.Attempts have been made to retrieve the device.To date the device has not been returned.If the device or further details are received at a later date a supplemental medwatch will be sent.A manufacturing record evaluation was performed for the finished device lot number r4177l, and no non-conformances were identified.
 
Event Description
It was reported that a hole was found in the tyvek of an er320.No patient involvement occurred.The customer believes that the sterility was compromised.
 
Manufacturer Narrative
(b)(4).Batch # r9536g.Device analysis: the analysis results found that an er320 device was returned inside its package unopened.Upon visual inspection, it was observed that the tyvek from the packaging was damaged; the damaged was confirmed to be wrinkled.The sterility was not compromised.Due to the damages found on the packaging, a possible cause for these conditions is due to improper handling during transit or storage.All ees product is 100% inspected prior to release.The information you provided is compiled monitored and reviewed on a routine basis for any associated trends.A manufacturing record evaluation was performed for the finished device batch number r9536g, and no non-conformances were identified.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Type of Device
CLIP, IMPLANTABLE
Manufacturer (Section D)
ETHICON ENDO-SURGERY, LLC.
475 calle c
guaynabo 00969
MDR Report Key8340040
MDR Text Key136797064
Report Number3005075853-2019-16614
Device Sequence Number1
Product Code FZP
UDI-Device Identifier10705036012580
UDI-Public10705036012580
Combination Product (y/n)N
PMA/PMN Number
K864102
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Type of Report Initial,Followup
Report Date 01/25/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date11/30/2023
Device Catalogue NumberER320
Device Lot NumberR4177L
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/26/2019
Initial Date Manufacturer Received 01/25/2019
Initial Date FDA Received02/14/2019
Supplement Dates Manufacturer Received02/26/2019
Supplement Dates FDA Received03/18/2019
Patient Sequence Number1
-
-