Device evaluation: one smiths medical cadd legacy pca pump was returned for analysis with no damage.Event log history was performed and showed that three 20ml bolus tests may have been performed prior to the pumps return to smiths medical.During analysis, the customer's concern regarding failed accuracy was able to be confirmed.The delivery accuracy tests found the pump to be over delivering outside the published specification of +/-6%.The pump's expulsor will be replaced in order to bring the delivery into more nominal of a range.Based on the evidence, the complaint was confirmed, and the problem source of the reported event is unknown.
|