MAUDE Adverse Event Report: DEPUY SYNTHES PRODUCTS LLC BATTERY REAMER/DRILL FOR BPL II; MOTOR, SURGICAL INSTRUMENT, AC-POWERED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY SYNTHES PRODUCTS LLC BATTERY REAMER/DRILL FOR BPL II; MOTOR, SURGICAL INSTRUMENT, AC-POWERED Back to Search Results
Catalog Number 530.705
Device Problem Intermittent Energy Output (4025)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Type  malfunction  
Manufacturer Narrative
Udi: (b)(4).As of this date, the device has not been returned for evaluation; therefore, the reported condition cannot be confirmed and/or duplicated.If additional information should become available, a supplemental medwatch report will be sent accordingly.
 
Event Description
It was reported that during an unspecified surgical procedure, it was observed that the battery reamer device worked intermittently.It was unknown if there were delays in the surgical procedure or if a spare device was available for use.There was patient involvement.There were no reports of injuries, medical intervention or prolonged hospitalization.The date of event was unknown.All available information has been disclosed.If additional information should become available, a supplemental medwatch will be submitted accordingly.
 
Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.The actual device was returned for evaluation.The device was evaluated and it was determined that the battery reamer device had no power on bench test, the motor and gearbox were corroded, and the gearbox rotation was not turning smoothly.Therefore, the reported condition was confirmed.The assignable root cause was determined to be due to improper maintenance.If additional information should become available, a suplemental medwatch will be submitted accordingly.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BATTERY REAMER/DRILL FOR BPL II
Type of Device
MOTOR, SURGICAL INSTRUMENT, AC-POWERED
Manufacturer (Section D)
DEPUY SYNTHES PRODUCTS LLC
eimattstrasse 3
oberdorf 4436
SZ  4436
MDR Report Key8343185
MDR Text Key136887373
Report Number8030965-2019-60992
Device Sequence Number1
Product Code GEY
UDI-Device Identifier10886982240125
UDI-Public10886982240125
Combination Product (y/n)N
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type user facility
Type of Report Initial,Followup
Report Date 02/05/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/15/2019
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number530.705
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/11/2019
Date Manufacturer Received02/27/2019
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-