• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITH & NEPHEW ORTHOPAEDICS LTD ACETABULAR CUP HAP SIZE 48/54; PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/METAL, RESURFACING

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW ORTHOPAEDICS LTD ACETABULAR CUP HAP SIZE 48/54; PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/METAL, RESURFACING Back to Search Results
Catalog Number 74122154
Device Problems Inaccurate Delivery (2339); Adverse Event Without Identified Device or Use Problem (2993)
Patient Problems Inflammation (1932); Pain (1994); Injury (2348)
Event Date 11/25/2013
Event Type  Injury  
Event Description
It was reported that a revision surgery from the left hip was performed due to persistent hip and groin pain and acute inflammation.
 
Manufacturer Narrative
It was reported that left hip revision surgery was performed.During the revision the acetabular cup & femoral head were removed.As of today, the implanted devices, all of which were used in treatment, and additional information has been requested for this complaint but has not become available.A review of the complaint history for the femoral head and the acetabular cup was performed using batch numbers in search of similar recurring reports for the products during their lifetimes.No other similar complaints were identified for the acetabular cup and the femoral head.In the absence of the actual devices, the production records were reviewed for the devices reportedly involved in this incident.Device history record review confirmed that all released parts met specifications applicable at the time of production.Review of the product ifu found adequate warnings and precautions in relation to the alleged failure modes.A risk management review was performed.No additional risks were identified as result of the reported event.The medical documents were reviewed.Although it was reported by the patient that there was concern about elevated cobalt and chromium serum levels and acute inflammation, neither the levels nor the lab reports were provided for review.Without the supporting lab results, imaging, and the analysis of the explanted components, the root cause of the reported pain, elevated metal ions and acute inflammation cannot be confirmed and it cannot be concluded that the reported clinical reactions were associated with a mal-performance of the implant.The patient impact beyond the pain, revision, and expected transient post-op convalescence period cannot be determined.Without return of the actual devices or further information we cannot further investigate or confirm the details supplied in this complaint, and our investigation remains inconclusive.If the products or additional information become available in the future, this case will be reopened.No preventative or corrective action has been initiated as a result of this investigation.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ACETABULAR CUP HAP SIZE 48/54
Type of Device
PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/METAL, RESURFACING
Manufacturer (Section D)
SMITH & NEPHEW ORTHOPAEDICS LTD
aurora house
spa park
leamington spa CV31 3HL
UK  CV31 3HL
MDR Report Key8345487
MDR Text Key136340851
Report Number3005975929-2019-00069
Device Sequence Number1
Product Code NXT
Combination Product (y/n)N
PMA/PMN Number
P040033
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type other
Type of Report Initial,Followup
Report Date 03/19/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Expiration Date12/31/2014
Device Catalogue Number74122154
Device Lot Number09MW25376
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 02/12/2019
Initial Date FDA Received02/15/2019
Supplement Dates Manufacturer Received02/12/2019
Supplement Dates FDA Received03/19/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Treatment
74123148 FEMORAL HEAD, LOT 10AW25585.; UNKNOWN BHR DEVICE/ UNKNOWN LOT.; UNKNOWN BHR DEVICE/ UNKNOWN LOT
Patient Outcome(s) Hospitalization; Required Intervention;
Patient Age47 YR
-
-