• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: AGA MEDICAL CORPORATION AMPLATZER SEPTAL OCCLUDER; TRANSCATHETER SEPTAL OCCLUDER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

AGA MEDICAL CORPORATION AMPLATZER SEPTAL OCCLUDER; TRANSCATHETER SEPTAL OCCLUDER Back to Search Results
Model Number 9-ASD-020
Device Problem Difficult or Delayed Positioning (1157)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 01/17/2019
Event Type  Injury  
Event Description
On (b)(6) 2019, a 18mm amplatzer sizing balloon (sb) and 20mm amplatzer atrial septal occluder (aso) were selected for implant.During device preparation, the 20mm sb would not deflate.During implant, the 20mm aso would not deploy properly and adjust properly to the patient's anatomy's.The procedure was abandoned due to unavailable stock and open surgery has been planned.The patient remained hemodynamically stable throughout the procedure and was reported with no adverse consequences.
 
Manufacturer Narrative
The reported event of device deformation could not be confirmed.The photo received from the field could not be analyzed and the investigation confirmed the device met visual and functional specifications when analyzed at abbott.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed and the product met all specifications at the time of commercialization.The cause of the reported incident could not be conclusively determined.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AMPLATZER SEPTAL OCCLUDER
Type of Device
TRANSCATHETER SEPTAL OCCLUDER
Manufacturer (Section D)
AGA MEDICAL CORPORATION
5050 nathan lane north
plymouth MN 55442
Manufacturer (Section G)
AGA MEDICAL CORPORATION
5050 nathan lane north
plymouth MN 55442
Manufacturer Contact
stephanie o' sullivan
5050 nathan lane north
plymouth, MN 55442
6517565400
MDR Report Key8346131
MDR Text Key136493772
Report Number2135147-2019-00040
Device Sequence Number1
Product Code MLV
UDI-Device Identifier00811806010175
UDI-Public00811806010175
Combination Product (y/n)N
Reporter Country CodeCO
PMA/PMN Number
P000039
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 04/10/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/15/2019
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Expiration Date12/31/2022
Device Model Number9-ASD-020
Device Catalogue Number9-ASD-020
Device Lot Number6296808
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/20/2019
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received04/08/2019
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured01/29/2018
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Hospitalization; Required Intervention;
-
-