ADVANCED BIONICS, LLC HIRES 90K¿ IMPLANT; COCHLEAR IMPLANT Back to Search Results
Model Number CI-1400-01
Device Problem Device Slipped (1584)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 01/24/2019
Event Type  malfunction  
Event Description
The recipient is reportedly experiencing loss of lock due to a head trauma incident.External equipment was exchanged and programming adjustments were made, however the issue did not resolve.Revision surgery is scheduled.
 
Manufacturer Narrative
The recipient's device was reportedly explanted.The recipient was reimplanted with another advanced bionics cochlear device.
 
Manufacturer Narrative
Advanced bionics considers the investigation into this reportable event as closed.The recipient's activation went well.The external visual inspection revealed the electrode was severed near the array and surgical tool marks were observed on the top cover prior to receipt.These are believed to have occurred during revision surgery.In addition, antenna coil damage was observed.The photographic imaging inspection confirmed antenna coil breaks.System lock could not be obtained at any spacing.The no lock condition prevented some of the electrical tests from being performed.The device passed some of the electrical tests performed.The device passed the mechanical tests performed.This device had broken antenna coil wires, which are consistent with the head trauma reported.A corrective action was implemented.This is the final report.Disclaimer: advanced bionics does not intend that this report be any admission of liability, fault or product defect.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HIRES 90K¿ IMPLANT
Type of Device
COCHLEAR IMPLANT
Manufacturer (Section D)
ADVANCED BIONICS, LLC
28515 westinghouse place
valencia CA 91355
MDR Report Key8347603
MDR Text Key136605954
Report Number3006556115-2019-00047
Device Sequence Number1
Product Code MCM
Combination Product (y/n)N
PMA/PMN Number
P960058
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,health professional
Type of Report Initial,Followup,Followup
Report Date 01/24/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Expiration Date09/30/2014
Device Model NumberCI-1400-01
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/15/2019
Initial Date Manufacturer Received Not provided
Initial Date FDA Received02/18/2019
Supplement Dates Manufacturer ReceivedNot provided
Not provided
Supplement Dates FDA Received03/18/2019
04/30/2019
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Age6 YR
-
-