Advanced bionics considers the investigation into this reportable event as closed.The recipient's activation went well.The external visual inspection revealed the electrode was severed near the array and surgical tool marks were observed on the top cover prior to receipt.These are believed to have occurred during revision surgery.In addition, antenna coil damage was observed.The photographic imaging inspection confirmed antenna coil breaks.System lock could not be obtained at any spacing.The no lock condition prevented some of the electrical tests from being performed.The device passed some of the electrical tests performed.The device passed the mechanical tests performed.This device had broken antenna coil wires, which are consistent with the head trauma reported.A corrective action was implemented.This is the final report.Disclaimer: advanced bionics does not intend that this report be any admission of liability, fault or product defect.
|