The customer reported the 130344, single function abc handpiece, caught fire at the tip during a laparoscopic cholecystectomy on "(b)(6)" 2018.The fire was in the sterile field at the time, but it was stomped out by the surgeon.There were no patient or user injuries.The surgical team was familiar with the product due to continuous education by nursing staff.The abc handpiece was stored in the holster in the sterile field when not in use.The handpiece was attached to the 7550 argon generator.The handpiece was checked prior to surgery.It is unknown how long the handpiece was in use before it caught fire.This report is being raised based on device malfunction with potential for injury upon reoccurrence.
|
Correction: previously stated catalog # 130344, lot # 201810294 corrected to catalog # 160656, no lot # provided.Previously stated k871435 corrected to k925903.Two 160656 were received, one in opened unoriginal packaging the other in opened original packaging, the reported catalog number was verified, the lot number was not verified.Visual inspection found that the device that did not have original packaging appeared burnt at the distal end and that the shaft and coating were damaged.The device was also functionally inspected with a system 7500 esu ((b)(4)).Functional inspection found that the argon beam ejected straight from the tip and did not appear to arc or creep along the sides.No smoke, fire, or sparking was observed during testing.The manufacturing documents from the device history record could not be reviewed due to no lot number being provided.(b)(4).Per the instructions for use, the user is advised the following; not to use in the presence of flammable anesthetics, disinfecting agents, oxygen-rich environments or other combustible materials.Before use, inspect the cord insulation and hand piece integrity and condition.If damaged, chipped, cut, or nicked, do not use the hand piece /probe.This issue will continue to be monitored through the complaint system to assure patient safety.
|