• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITH & NEPHEW ORTHOPAEDICS LTD UNKNOWN BHR DEVICE; PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/METAL, RESURFACING

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW ORTHOPAEDICS LTD UNKNOWN BHR DEVICE; PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/METAL, RESURFACING Back to Search Results
Device Problems Inadequacy of Device Shape and/or Size (1583); Noise, Audible (3273); Patient Device Interaction Problem (4001)
Patient Problems Hypersensitivity/Allergic reaction (1907); Pain (1994); Injury (2348)
Event Date 02/06/2017
Event Type  Injury  
Event Description
It was reported that a revision surgery from the right hip was performed due to allergic reaction to the metal-on- metal rubbing, metal debris, popping, pain, grinding and different leg lengths.
 
Manufacturer Narrative
It was reported that right hip revision surgery was performed.During the revision the hemi head & modular sleeve were removed.The acetabular cup & stem remained implanted.As of today, the implanted devices, all of which were used in treatment, and additional information has been requested for this complaint but has not become available.Without definitive part/lot numbers a complete complaint history review cannot be performed for the devices involved.A review of the complaint history review was performed using the part number for a 74122550 hemi head 50mm, 74120150 acetabular cup 50mm & 74222300 modular sleeve +4mm.Similar complaints have been identified and this failure will continue to be monitored.Without the details of the devices involved in this complaint, the specific product labelling and ifus for the devices cannot be reviewed.A risk management review was performed.No additional risks were identified as result of the reported event.The medical documents were reviewed.It should be noted bhr is contraindicated in patients with known or suspected metal sensitivity.It cannot be determined to what extent the patient¿s fractured ilium and displaced fragment and advanced facet arthropathy had on her pain and clinical status.Cannot confirm leg-length discrepancy with the documentation provided.Pain, inflammation and adverse tissue reaction are findings associated with an allergic response.However, it cannot be concluded that the reported clinical reactions were associated with a mal-performance of the implant.The patient impact beyond the pain, revision, and expected transient post-op convalescence period cannot be determined.Without return of the actual devices or further information we cannot further investigate or confirm the details supplied in this complaint, and our investigation remains inconclusive.If the products or additional information become available in the future, this case will be reopened.No preventative or corrective action has been initiated as a result of this investigation.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
UNKNOWN BHR DEVICE
Type of Device
PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/METAL, RESURFACING
Manufacturer (Section D)
SMITH & NEPHEW ORTHOPAEDICS LTD
aurora house
spa park
leamington spa CV31 3HL
UK  CV31 3HL
MDR Report Key8352610
MDR Text Key136580792
Report Number3005975929-2019-00076
Device Sequence Number1
Product Code NXT
Combination Product (y/n)N
PMA/PMN Number
P040033
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type other
Remedial Action Recall
Type of Report Initial,Followup
Report Date 03/03/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 02/12/2019
Initial Date FDA Received02/19/2019
Supplement Dates Manufacturer Received02/12/2019
Supplement Dates FDA Received03/03/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Treatment
SMITH & NEPHEW ANTHOLOGY STEM; UNKNOWN BHR DEVICE/ UNKNOWN LOT
Patient Outcome(s) Hospitalization; Required Intervention;
-
-