MAUDE Adverse Event Report: BIOMET FRANCE S.A.R.L. OPTIPAC 80 REFOBACIN BONE CEMENT R-3; BONE CEMENT, ANTIBIOTIC

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BIOMET FRANCE S.A.R.L. OPTIPAC 80 REFOBACIN BONE CEMENT R-3; BONE CEMENT, ANTIBIOTIC Back to Search Results
Catalog Number 4712500398-3
Device Problem Infusion or Flow Problem (2964)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 04/10/2018
Event Type  malfunction  
Manufacturer Narrative
(b)(4).The device was not returned to the manufacturer.Therefore it could not be analyzed.The device manufacturing quality record indicates that the released product met all requirements to perform as intended.Investigation results concluded that the product is conform and the root cause is undetermined.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly zimmer biomet will continue to monitor for trends.
 
Event Description
It has been reported that the monomer fluid wasn't injected into the cartridge after having pressed the blue brackets.
 
Event Description
It has been reported that the monomer fluid wasn't injected into the cartridge after having pressed the blue brackets.
 
Manufacturer Narrative
(b)(4).This follow-up report is being submitted to relay additional information.The device manufacturing quality record indicates that the released product met all requirements to perform as intended.Complaint sample was evaluated and the reported event was confirmed.No product non-conformity was identified.Product analysis showed that the most probable root cause is a handling error.Investigation results concluded that the reported event was likely due to an handling error.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
OPTIPAC 80 REFOBACIN BONE CEMENT R-3
Type of Device
BONE CEMENT, ANTIBIOTIC
Manufacturer (Section D)
BIOMET FRANCE S.A.R.L.
plateau de lautagne bp75
valence cedex 26903
FR  26903
MDR Report Key8357458
MDR Text Key136786383
Report Number3006946279-2019-00101
Device Sequence Number1
Product Code MBB
UDI-Device Identifier04040029922385
UDI-Public(01)04040029922385
Combination Product (y/n)N
PMA/PMN Number
K171540
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign
Type of Report Initial,Followup
Report Date 05/16/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date11/30/2018
Device Catalogue Number4712500398-3
Device Lot Number712BA07300
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/29/2018
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 04/12/2018
Initial Date FDA Received02/21/2019
Supplement Dates Manufacturer Received05/15/2019
Supplement Dates FDA Received05/17/2019
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-