Catalog Number 257618110
Device Problems Material Discolored (1170); Naturally Worn (2988); Scratched Material (3020)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 01/31/2019
Event Type  malfunction  
Manufacturer Narrative
If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.Device is an instrument and is not implanted/explanted.
 
Event Description
It was reported that instruments- srom sleeve trial was worn and can't read sizing info on them.16d small; 18b small; 16d large.Broken drill bit.Broke on cutting flutes.25mm.Drill bit had been reused multiple times.No surgical delay.
 
Manufacturer Narrative
Product complaint # (b)(4).Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.Investigation summary : examination of the returned instrument confirmed the complaint.Depuy considers the investigation closed at this time.Should additional information be received, the information will be reviewed and the investigation will be re-opened as necessary.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ZT TRIAL SLEEVE 18D SML
Type of Device
HIP INSTRUMENTS : FEMORAL TRIALS
Manufacturer (Section D)
DEPUY ORTHOPAEDICS INC US
700 orthopaedic drive
warsaw IN 46581 0988
MDR Report Key8361718
MDR Text Key136865046
Report Number1818910-2019-85503
Device Sequence Number1
Product Code LXH
UDI-Device Identifier10603295146605
UDI-Public10603295146605
Combination Product (y/n)N
PMA/PMN Number
UNK
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Type of Report Initial,Followup
Report Date 01/31/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number257618110
Device Lot NumberMT44299
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/18/2019
Initial Date Manufacturer Received 01/31/2019
Initial Date FDA Received02/22/2019
Supplement Dates Manufacturer Received02/21/2019
Supplement Dates FDA Received03/20/2019
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-