• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: PENUMBRA, INC. PENUMBRA SYSTEM ASPIRATION PUMP MAX 110; JCX

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

PENUMBRA, INC. PENUMBRA SYSTEM ASPIRATION PUMP MAX 110; JCX Back to Search Results
Catalog Number PMX110
Device Problems Mechanical Problem (1384); Suction Problem (2170)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 01/24/2019
Event Type  malfunction  
Manufacturer Narrative
The device has been returned and the investigation results are pending.A follow up mdr will be submitted upon completion of the device investigation.
 
Event Description
During preparation for a thrombectomy procedure, the hospital staff noticed that the regulator knob of a penumbra system pump max 110v (pump max) was broken.Although the reported issue was found prior to use and prior to the patient being on the table, the pump max was used in the procedure.During the procedure, the pump max produced low vacuum.The procedure was completed using the same pump max.There was no report of an adverse effect to the patient.
 
Manufacturer Narrative
Results: the regulator knob was loose from the pump housing.Conclusion: evaluation of the returned pump max revealed that the regulator knob was loose from the pump housing.If the regulator knob is over-tightened during use, the regulator knob may become loose.The loose regulator knob may contribute to difficulty regulating vacuum pressure.During functional testing, the regulator knob was able to regulate vacuum pressure.Penumbra pumps are visually inspected and functionally tested during incoming inspection by quality.The manufacturing records for this lot were reviewed and did not reveal any outstanding discrepancies, design, or quality concerns.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PENUMBRA SYSTEM ASPIRATION PUMP MAX 110
Type of Device
JCX
Manufacturer (Section D)
PENUMBRA, INC.
one penumbra place
alameda CA 94502
MDR Report Key8364563
MDR Text Key136995430
Report Number3005168196-2019-00344
Device Sequence Number1
Product Code JCX
UDI-Device Identifier00814548012773
UDI-Public00814548012773
Combination Product (y/n)Y
PMA/PMN Number
K122756
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Type of Report Initial,Followup
Report Date 01/01/2005,01/24/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberPMX110
Device Lot NumberF26270-17
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/05/2019
Was the Report Sent to FDA? Yes
Date Report Sent to FDA01/01/2005
Date Report to Manufacturer01/10/2005
Initial Date Manufacturer Received 01/24/2019
Initial Date FDA Received02/22/2019
Supplement Dates Manufacturer Received04/02/2019
Supplement Dates FDA Received04/29/2019
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Age88 YR
-
-