• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ABBOTT DIABETES CARE INC FREESTYLE LIBRE; FLASH GLUCOSE MONITORING SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ABBOTT DIABETES CARE INC FREESTYLE LIBRE; FLASH GLUCOSE MONITORING SYSTEM Back to Search Results
Model Number 71702-01
Device Problem Appropriate Term/Code Not Available (3191)
Patient Problems Skin Irritation (2076); Burning Sensation (2146)
Event Date 01/08/2019
Event Type  Injury  
Manufacturer Narrative
The product has been requested back for an investigation.A follow-up report will be submitted once additional information is obtained.The device mfg date is unknown.All pertinent information available to abbott diabetes care has been submitted.
 
Event Description
Customer reported experiencing an adverse skin reaction while wearing an adc freestyle libre sensor.Customer experienced ¿burning open wounds¿ and had contact with a healthcare professional on (b)(6) 2019 who prescribed cortisone, valaciclovir (antiviral) and desloratadine (antihistamine) for treatment.There was no report of death or permanent injury associated with this event.There was no report of death or permanent injury associated with this event.
 
Manufacturer Narrative
No product has been returned.Extended investigation has been performed for the reported complaint and there was no indication that the product did not meet specification.The reported complaint is related to skin irritation or an allergic reaction to the patch adhesive of the freestyle libre sensor.Dose audit and environmental monitoring reports were reviewed for issues relating to sterility of the product.The sensor component has no impact on product sterility and therefore, sensor components dhrs were not reviewed.Sensor kit dhrs have been reviewed to assess the manufacturing process, which includes the application of the adhesive to the puck.A dhr (device history review) for the freestyle libre sensor kit was reviewed and the dhr showed the freestyle libre sensor kit passed all tests prior to release.Dose audit reports were reviewed and demonstrates the continued effectiveness of the established sterilization process for libre sensor kits.Environmental monitoring reports were reviewed, including bioburden and endotoxin testing, and demonstrated that all monitoring processes continue to meet adc minimum requirements for product quality.All review activities conducted above, including but not limited to the final release testing specifically associated with the manufacture of this product, are sufficient information in order to show if the product has met specifications prior to release.If the product is returned, the case will be re-opened and a physical investigation will be performed.
 
Event Description
Customer reported experiencing an adverse skin reaction while wearing an adc freestyle libre sensor.Customer experienced ¿burning open wounds¿ and had contact with a healthcare professional on (b)(6) 2019 who prescribed cortisone, valaciclovir (antiviral) and desloratadine (antihistamine) for treatment.There was no report of death or permanent injury associated with this event.There was no report of death or permanent injury associated with this event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FREESTYLE LIBRE
Type of Device
FLASH GLUCOSE MONITORING SYSTEM
Manufacturer (Section D)
ABBOTT DIABETES CARE INC
1360 south loop road
alameda CA 94502 7001
MDR Report Key8368569
MDR Text Key137117260
Report Number2954323-2019-01651
Device Sequence Number1
Product Code PZE
Combination Product (y/n)N
PMA/PMN Number
P160030
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type consumer
Type of Report Initial,Followup
Report Date 06/18/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/25/2019
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Lay User/Patient
Device Model Number71702-01
Was Device Available for Evaluation? No
Date Manufacturer Received05/28/2019
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient Age56 YR
Patient Weight58
-
-