SMITH & NEPHEW, INC. PATELLA RESECTION GUIDE; PROSTHESIS, KNEE, FEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, METAL/POLYMER
|
Back to Search Results |
|
Catalog Number 71440391 |
Device Problem
Loose or Intermittent Connection (1371)
|
Patient Problem
No Consequences Or Impact To Patient (2199)
|
Event Date 02/04/2019 |
Event Type
malfunction
|
Event Description
|
It was reported that a screw came loose during procedure and fell on patient.All pieces accounted for.No delay or injury reported.A back up device was available.
|
|
Manufacturer Narrative
|
The associated device was returned and evaluated.A visual inspection of the returned device confirmed the stated failure mode.One of the screws used to assemble the locator arm came out, rendering the device inoperable.The screw was returned.The device was manufactured in 2017 and exhibits signs of moderate wear/usage.A review of complaint history did not reveal additional complaints for the listed batch.At this time, we have no reason to suspect that the product failed to meet any product specifications at the time of manufacture.We recommend that all reusable instruments be routinely inspected for wear and damage and replaced as necessary.Should additional information be received, the complaint will be reopened.No further investigation is warranted for this complaint; however we will continue to monitor for future complaints and investigate as necessary.
|
|