A getinge service technician has been sent over for investigation on 2019-03-11, where the failure was serviced and detected.The vacuum valve and the main board were replaced.The technician confirmed that functional tests has been performed and found ok.The device is back in clinical use.A life cycle engineering (lce) investigation was performed in order to determine the root cause of the reported issue.According to lce investigation report# (b)(4) following was found: the lce performed visual,electrical and mechanical tests.It was complained, that the vacuum valve was leaking.This could be confirmed, but not reproduced.A cause for the leakage could not be identified.Additionally, it was mentioned that the device board had been destroyed by a fluid.This can be confirmed, too.It is plausible that the fluid caused a short circuit on the pcb, which then triggered the complained burning smell and the fuse.Unusually, it was noticed that the drip protection at the housing of the pcb was not present.Reason not found.Thus the failure could be confirmed.The device was part of fsca-2018-07-18, in which the vacuum valve was replaced on 2018-11-26.This supports the above findings of the technician, that the leaking of the valve could be confirmed but not reproduced.The most probable root cause could be determined by fluid that got on the device board.There was a short circuit on the board, which destroyed the board.This also explains the complaint about the smell of fire and the triggering of the fuse.Since the reported failure did not contribute to a death or serious injury no corrective action is needed.In addition at this time it cannot be concluded that this is a systemic error.No corrective action is needed.Maquet medical systems,usa(importer)submits this report on behalf of the legal manufacturer of the device maquet cardiopulmonary ag kehler strasse 31, 76437 rastatt, germany.A follow-up medwatch will be submitted when additional information becomes available.Reference exemption #: e2018002.Importer: maquet medical systems usa 45 barbour pond drive wayne, nj 07470.Contact person: (b)(4).
|