MEDTRONIC PUERTO RICO OPERATIONS CO. ACTIVA; STIMULATOR, ELECTRICAL, IMPLANTED, FOR PARKINSONIAN TREMOR
|
Back to Search Results |
|
Model Number 37602 |
Device Problems
Shipping Damage or Problem (1570); Connection Problem (2900); Impedance Problem (2950)
|
Patient Problem
No Known Impact Or Consequence To Patient (2692)
|
Event Date 02/06/2019 |
Event Type
malfunction
|
Manufacturer Narrative
|
Other applicable components are: product id: neu_set_screw, serial# unknown, product type: accessory.If information is provided in the future, a supplemental report will be issued.
|
|
Event Description
|
Information was received from a healthcare provider (hcp) via a manufacturing representative (rep) regarding a patient with an implantable neurostimulator (ins) for parkinson's disease.It was reported that during the replacement of the ins on both sides, the screw of the #2 electrode of the ins on the left side, which was newly unpacked, could not be tightened and the impedance of the #2 electrode was abnormal.The impedance was checked several times after disconnection and reconnections, however no improvement was seen.A new screw was unpacked and used, and the procedure continued as the impedance was normal.The event was considered resolved.This was suspected to be an out of box issue.No patient symptoms or further complications were reported as a result of this event.
|
|
Manufacturer Narrative
|
If information is provided in the future, a supplemental report will be issued.
|
|
Event Description
|
Additional information received from a rep indicated the cause of the screw not tightening was not known.
|
|
Manufacturer Narrative
|
If information is provided in the future, a supplemental report will be issued.
|
|
Event Description
|
Additional information received from a rep indicated the ins was originally considered to have an anomaly on circuit #2.
|
|
Manufacturer Narrative
|
Analysis of the ins (serial no.(b)(4)) was performed, despite there being no original allegation on the device.The ins was returned with 4 set screws, and no anomaly was found with the set screws, however the repair analyst noted that the problem set screw was likely not returned.Correspondence with the manufacturing representative indicated the location of this set screw was unknown.The ins was found to have no malfunction.Product analysis #241045567:analysis information -- 2019-04-05 13:34:34 cst pli# 10 product id# 37602 below is unedited, system generated text based on the analysis finding code(s).The returned device was subjected to a series of standard tests that include but is not limited to visual inspection, output and telemetry testing, and functional testing.The implantable neurostimulator (ins) passed functional testing.If information is provided in the future, a supplemental report will be issued.
|
|
Search Alerts/Recalls
|
|
|